You have to be registered and logged in for purchasing articles.

Abstract

Establishing Conditions for Blood Smear Drying and Staining on Sysmex SP-50 for Leukocyte Differential Count by Hyun Lee, Jihye Kim, Jiwon Lee, Jisoo Won, Yoon Hwan Chang

Background: The purpose of this study was to determine the staining conditions and appropriate fan1 start time (FAN1ST) for Sysmex SP-50 to produce blood smears (BS) that reflect the true lymphocyte morphology of patient samples.
Methods: Using different start times of fan1, we obtained a set of 84 blood smear slides from 21 blood samples and measured 10,920 lymphocyte areas, which were then converted to compare lymphocyte sizes. We also performed a leukocyte differential count using Sysmex DI-60 on 202 blood smear slides prepared before and after the change in staining conditions and compared the results.
Results: The mean lymphocyte sizes at FAN1ST 0 second, 5 seconds, 10 seconds, and 30 seconds were 12.55 μm, 12.14 μm, 11.27 μm, and 10.50 μm, respectively. The mean differences in the preclassification of neutrophils, lymphocytes, monocytes, eosinophils, and basophils in DI-60, according to the SP-50 staining conditions, were 0.88, -1.58, -0.24, 0.37, and 0.07, respectively.
Conclusions: Wright-Giemsa staining of blood smears prepared on the SP-50 showed that changing the pH of the concentrated phosphate buffer to 6.6 and adjusting the staining time did not affect the results of the leukocyte differential count. However, since fan1 was used to dry the blood smear on the SP-50 and the lymphocyte size gradually decreased as the start time was delayed, it was necessary to set a start time for fan1 that did not affect the lymphocyte size.

DOI: 10.7754/Clin.Lab.2023.231029