Background: Hemoglobin A1c (HbA1c) is a widely accepted marker for long-term glycemic control. However, the presence of hemoglobin variants can interfere with its measurement, depending on the analytical method used. Accidental detection of silent variants during HbA1c monitoring is a known challenge in clinical laboratories.
Methods: A 79-year-old male patient with no known history of diabetes or hemoglobinopathy was admitted for chest pain. Routine laboratory tests were performed. HbA1c measurement was initially requested and performed using capillary electrophoresis (Capillarys OCTA 3, Sebia) in HbA1c mode. Confirmatory investigations included hemoglobin electrophoresis in both alkaline and acidic media, as well as High-Performance Liquid Chromatogra-phy (HPLC) using the ADAMS A1c HA-8180V (Arkray).
Results: Routine biochemical and hematological parameters were within normal ranges. The initial capillary electrophoresis analysis yielded an atypical electrophoretic profile characterized by split peaks for both HbA0 and HbA1c fractions. A minor abnormal fraction (2.3%) was also detected. Due to this interference, no HbA1c value could be reported. In contrast, further analysis using HPLC and standard gel electrophoresis returned normal results, suggesting the presence of a variant detectable specifically under the separation conditions of the capillary method used.
Conclusions: This case illustrates the potential for analytical interference by a rare hemoglobin variant, which altered the electrophoretic migration of HbA0 and HbA1c in capillary electrophoresis. The discrepancy between techniques highlights the importance of using complementary methods to ensure diagnostic accuracy when an atypical profile is observed. In such situations, alternative markers like fructosamine or glycated albumin should be considered for glycemic monitoring.
DOI: 10.7754/Clin.Lab.2025.251223
|