Comparison of Two HbA1c Measurement Methods and Their Analytical Performance: TINIA vs. Boronate Affinity HPLC by Nermin Dindar-Badem, Seydi A. Peker, Yasemin T. Sutaşir

Background: Hemoglobin A1c (HbA1c) measurement is essential for the assessment of long-term glycemic control in patients with diabetes mellitus, as it reflects the average blood glucose concentration over the preceding 8 - 12 weeks. Global diabetes mellitus prevalence has increased dramatically, and HbA1c testing is critical for diagnosis, monitoring treatment efficacy, evaluating diabetes care quality, and predicting the risk of diabetic complications. Although HbA1c has become the standard method for assessing long-term glycemic control in diabetic patients, various measurement methods exist with different analytical characteristics and potential interferences. This study aimed to evaluate and compare the analytical performance of the Roche turbidimetric inhibition Immunological assay (TINIA) method with the Premier boronate affinity high-performance liquid chromatography (HPLC) method for HbA1c measurement.
Methods: Blood samples were randomly selected from 75 diabetic and non-diabetic patients (48 females [64%], 27 males [36%], median age 55 years). HbA1c levels were measured using both Roche TINIA (cobas c501 analyzer) and Premier Hb9210 boronate affinity HPLC methods. Analytical performance was evaluated through precision studies, linearity analysis, bias estimation, and comparison studies using Passing-Bablok regression and Bland-Altman analyses according to CLSI EP-5 protocol.
Results: Both methods demonstrated acceptable precision with coefficient of variation (CV%) values below 2.0%, meeting the IFCC-recommended criteria of < 2.8%. The correlation between the two methods was strong (r = 0.986, p < 0.0001), with mean HbA1c values of 7.16 ± 2.14% for Roche TINIA and 6.99 ± 1.94% for Premier Hb9210. Despite excellent correlation, a statistically significant difference was found between methods (p = 0.0005). The Bland-Altman analysis showed a mean difference of -0.17% (95% CI: -1.9 to 1.5%) between methods, with mean relative bias of 0.17%, which was well below the IFCC-defined total allowable error criterion of 0.46% at 6.7% HbA1c levels.
Conclusions: Both Roche TINIA and Premier Hb9210 boronate affinity HPLC methods demonstrated acceptable analytical performance and excellent agreement for HbA1c measurement. The TINIA method showed good concordance with the reference boronate affinity chromatography method, supporting its clinical utility for diabetes management. However, clinicians should be aware of the small but consistent bias between methods when monitoring patients over time or when switching analytical platforms.

DOI: 10.7754/Clin.Lab.2025.251123